Traditional QCM technology is limited to characterization of thin and rigid films. This limitation, and the possibility to characterize soft or thick films has been addressed with extended QCMs, such as QCM-D, which also measure the energy losses. Here we explain how QCM-D works.
Quartz crystal microbalance with Dissipation,QCM-D, is based on the sameworking principleasQCM, i.e. it measures the resonance frequency changes,Δf, of a quartz crystal as a function of time. In addition to the frequency capture, QCM-D also collects time resolved information about the energy losses, thedissipation. I. e. QCM-D collectsΔfandΔDas a function of time, Fig 1, whereas QCM collectsΔfas a function of time.
Figure 1. The figure shows a schematic illustration of the working principle of QSense QCM-D, a multi-harmonic QCM-D, which via so-called pinging provides time resolved information on f and D at multiple harmonics.Via a rapid, periodic excitation of the driving power (upper left), the oscillation frequency, f, and the decay time, which is inversely proportional to D, can be captured and thereby, time-resolved information of changes inΔf andΔD can be plotted as molecules interact with the sensor surface. In the schematic adsorption scenario and plot, section (I) shows a bare surface and stable baselines of multipleΔf:s andΔD:s. Period (II) shows how molecules bind to the surface, and as a result, the frequencies decrease and the dissipations increase, indicating mass uptake and increasing energy loss. Period (III) shows how the surface uptake has been completed and the frequency and dissipation responses have stabilized. The collectyed QCM-D data can be used for viscoelastic modelling and quantification of mass, thickness and viscoelastic properties of the adsorbed layer.
The benefit of measuring the dissipation in addition to just the frequency, is that you get information on whether the layer at the sensor surface is rigid or not, and if the Sauerbrey equation can be used for the quantification of mass. Also, iffandDare collected at multiple harmonics, it is possible to apply viscoelastic modelling on the data.
QCM-D technology measures two parameters in parallel, the resonance frequency,f, and the dissipation,D.
QCM-D:s currently in the market measure the dissipation either via the bandwidth, or via measuring the oscillation decay time after the driving power has been switched off, so-called pinging. QSense QCM-D is based on the pinging approach.
Download the overview to learn more about QCM-D technology and how it works.
Learn about what aspects to consider when setting up your QCM measurement
Learn more about QSense 4th generation QCM-D platform
Temperature stability is critical for reliable QCM data. Here are the top four factors that will help you eliminate temperature induced artifacts.
Sensitivity vs Limit of detection (LOD) - read about how they compare and why one is more imoprtant than the other.
Learn about lipid nanoparticle designs for drug delivery purposes and related research
Read about the key signatures in the QCM data that reveal if viscoelastic modelling should be used to extract the mass.
Viscoelasticity is a quality involving both viscous and elastic properties at the same time.
读到不同的药物参数是什么意思and which ones you should keep an eye on
QCM-sensor regeneration can be a resource efficient way of running experiments. Learn about aspects to consider if you plan to resuse ypur sensors.
Read about how QSense QCM-D analysis was used to better understand why and what affects an increased transmissibility of SARS-CoV-2